GRAS RA0045-S1 Prepolarized Ear Simulator According to IEC 60318-4 (60711)

  •  Part Number: GRAS RA0045-S1
     Category: GRAS Sound & Vibration
  •  View: 12
     Status: Available
  • Volume: 1260 mm³ @ 500 Hz
    Dyn range: 25 dB(A) to 150 dB
    Sensitivity 12.5 mV/Pa
    IEC: 60318-4

  • Sale Price: Liên hệ
Product Information

GRAS RA0045-S1 is an ear simulator with an acoustic input impedance closely resembling that of an average human ear. The RA0045 includes a 40AO 1/2" prepolarized pressure microphon and each RA0045 is individually calibrated with this specific microphone.

Technology

Introduction

The RA0045-S1 Prepolarized Ear Simulator is for making acoustic measurements on earphones coupled to the human ear by ear inserts such as tubes, ear moulds, or ear tips. The acoustic input impedance of RA0045-S1 closely resembles that of the human ear and, as a result, loads a sound source in very much the same way.

The RA0045-S1 complies with the following international requirements:

  • IEC 60318-4: Occluded-ear simulator for the measurement of earphones coupled to the ear by ear inserts.
  • ITU-T Recommendations P.57 (08/96) Series P: Telephone transmission quality, Objective measuringapparatus: Artificial ears.

It is measured and calibrated according to the ITU-T Recommendation P.57.

Design

The RA0045-S1 embodies a number of carefully designed volumes connected via well-defined and precisely tuned resistive grooves. In an equivalent electrical circuit, capacitors would represent the volumes, and inductance and resistance would represent respectively air mass and air flow within the resistive groves.

It is delivered with a built-in GRAS 40AO 1/2" prepolarized pressure microphone and an individual calibration chart for the ear simulator.

Typical Applications and Use

The RA0045-S1 is for measurements on earphones and hearing aids coupled to the human ear by ear inserts such as tubes, ear moulds, or ear tips.

In accordance with ITU-T Recommendation P.57, it can be used with the following GRAS pinna simulators for testing telephones:

  • GRAS RA0056 Low-leak Pinna Simulator 
  • GRAS RA0057 High-leak Pinna Simulator

It is also part of the GRAS 43AC-S1 Artificial Ear.

A comprehensive range of accessories for making measurements in accordance with IEC 60318-4 are available. See the tab Ordering info.

Compatibility

The RA0045-S1 can be used with a standard CCP preamplifier, e.g. the GRAS 26CA 1/2" Preamplifier or the GRAS 26CB 1/4" Preamplifier fitted with an adapter. For the 1/4" preamplifier, use either the straight RA0003 Adapter or the right-angled RA0001 Adapter.

Specifications

Polarization/Connection 0 V / CCP
Lower limit of theoretical dynamic range with GRAS preamplifier dB(A) 25
Upper limit of dynamic range with GRAS CCP preamplifier dB 150
Sensitivity set at 250 Hz (±3 dB) mV/Pa 12.5
Sensitivity set at 250 Hz (±3 dB) dB re 1V/Pa -38
Resonance frequency kHz 13.5
Coupler volume mm³ 1260 @ 500 Hz
Operating temperature range °C / °F -30 to 60 / -22 to 140
Temperature coefficient @250 Hz dB/°C / dB/°F 0.05
Non-condensing humidity range % RH 0 to 90
ANSI standard S3.7
IEC standard 60318-4 (formerly 60711)
ITU-T recommendation Page 57
CE/RoHS/WEEE compliant Yes/Yes/Yes
Weight g/oz 52 / 1.8

Typical frequency response

GRAS Sound & Vibration reserves the right to change specifications and accessories without notice.

Dimensions