The GRAS 90CA-S2 Microphone Calibration System provides for automated and reliable level and frequency response calibration of measurement microphones. The implemented calibration methods conform with these standards:
- IEC Standard 61094
- ANSI S1.10 – 1966 (R1976)
- IEC Standard 60942
- TEDS Standard IEEE 1451.4
The 90CA-S2 is based on National Instruments\' PXI architecture: NI PXIe-8101 Embedded Controller, including: National Instruments PXIe1071 4-Slot 3U PXI Express Chassis and PXI cards for Control and Data Acquisition.
Features
- Ready-to-use software and hardware
- Accurate calibration hardware
- Automated and computer-controlled calibration process
- Highly reproducible calibration method
- Pre-installed test parameters for GRAS microphones
- Test conditions easily changable
- Auto-generated, customized documentation to print and archive
Applications
- All types of IEC 61094 standardized 1\", 1/2\", 1/4\", and 1/8\" measurement microphones from GRAS and other manufacturers
- GRAS 46xx series standard microphone sets
- GRAS 47xx series flush-mount microphone sets (optional)
- GRAS preamplifiers
- Calibration up to 92 kHz (optional up to 200 kHz)
The system is described in more detail in the Instruction Manual.
Specifications
| IEC Standard |
– |
61094-5 & 61094-6 |
| Sensitivity calibration – Method |
– |
Pistonphone, class 0 |
| Sensitivity calibration – Frequency |
Hz |
250 |
| Sensitivity calibration – Uncertainty |
dB |
0.07 |
| Frequency calibration – Method |
– |
Electrostatic actuator |
| Frequency calibration – Frequency |
Hz |
20 to 92 kHz (20 to 200 kHz)* |
| Frequency calibration – Uncertainty |
dB |
0.3 |
* Used with HP34410a multimeter (or equivalent), not included with the product.