GRAS 43AC-S6 Hi-Res Ear Simulator Kit LEMO

  •  Part Number: GRAS 43AC-S6
     Category: GRAS Sound & Vibration
  •  View: 15
     Status: Available
  • Freq range: 100 Hz to 50 kHz
    Dyn range: 44 dB(A) to 166 dB
    Sensitivity: 1.6 mV/Pa
    Use: For general acoustic diagnostics

  • Sale Price: Liên hệ
Product Information

The GRAS 43AC-S6 Ear Simulator Kit is a complete test-jig for hi-res measurements up to 50 kHz on earphones coupled to the ear by inserts such as tubes and earmoulds.

Technology

The GRAS 43AC-S6 Ear Simulator Kit is a complete test jig for acoustically testing earphones coupled to the ear via inserts such as tubes and ear moulds. It is compatible with the following international requirements:

  • IEC 60318-4: Occluded-ear simulator for the measurement of earphones coupled to the ear by ear inserts.

43AC-S6 uses the RA0403 Hi-Res Ear Simulator which extends the useful frequency range to 50 kHz. Read more about the RA0403 here.

The 43AC comprises:

  • GRAS RA0403 Externally Polarized Hi-Res Ear Simulator

  • GRAS 40BP 1/4" Pressure Microphone (included in RA0403)

  • GRAS 26AC 1/4" Preamplifier

  • GRAS RA0001 Right-angled Adapter

  • GRAS RA0052 Test Jig

The Test Jig has an adjustable spring-loaded arm to exert a variable force on the test object.

A prepolarized version is available,.

Specifications

 

Parameter Unit Value
Theoretical dynamic range lower limit with GRAS preamplifier dB(A) 44
Theoretical dynamic range upper limit with GRAS preamplifier @ +120 V / ±60 V power supply dB 169
Set sensitivity @ 250 Hz (±3 dB) mV/Pa 1.6
Coupler volume mm³ 1260 @ 500 Hz
Resonance frequency kHz 13.5
Temperature range, operation °C / °F -30 to 60 / -22 to 140
Temperature coefficient @250 Hz dB/°C / dB/°F -0.01 / -0.006
Humidity range non-condensing % RH 0 to 75
ANSI standard S3.7
IEC standard 60318-4 (former 60711)
CE/RoHS compliant / WEEE registered Yes / Yes / Yes
Connector type 3 m 7-pin LEMO
Weight g / oz 1550 / 54.675

 

Fig 1. The transfer impedance of a typical RA0403 Hi-Res Ear Simulator with tolerances. Up to 10 kHz, the response is within the limits defined by IEC 60318-4.  Above 10 kHz, the response is within a fairly tight tolerance band up to 50 kHz