The GRAS 43AC-S6 Ear Simulator Kit is a complete test-jig for hi-res measurements up to 50 kHz on earphones coupled to the ear by inserts such as tubes and earmoulds.
Technology
The GRAS 43AC-S6 Ear Simulator Kit is a complete test jig for acoustically testing earphones coupled to the ear via inserts such as tubes and ear moulds. It is compatible with the following international requirements:
43AC-S6 uses the RA0403 Hi-Res Ear Simulator which extends the useful frequency range to 50 kHz. Read more about the RA0403 here.
The 43AC comprises:
-
GRAS RA0403 Externally Polarized Hi-Res Ear Simulator
-
GRAS 40BP 1/4" Pressure Microphone (included in RA0403)
-
GRAS 26AC 1/4" Preamplifier
-
GRAS RA0001 Right-angled Adapter
-
GRAS RA0052 Test Jig
The Test Jig has an adjustable spring-loaded arm to exert a variable force on the test object.
A prepolarized version is available,.
Specifications
| Parameter |
Unit |
Value |
| Theoretical dynamic range lower limit with GRAS preamplifier |
dB(A) |
44 |
| Theoretical dynamic range upper limit with GRAS preamplifier @ +120 V / ±60 V power supply |
dB |
169 |
| Set sensitivity @ 250 Hz (±3 dB) |
mV/Pa |
1.6 |
| Coupler volume |
mm³ |
1260 @ 500 Hz |
| Resonance frequency |
kHz |
13.5 |
| Temperature range, operation |
°C / °F |
-30 to 60 / -22 to 140 |
| Temperature coefficient @250 Hz |
dB/°C / dB/°F |
-0.01 / -0.006 |
| Humidity range non-condensing |
% RH |
0 to 75 |
| ANSI standard |
– |
S3.7 |
| IEC standard |
– |
60318-4 (former 60711) |
| CE/RoHS compliant / WEEE registered |
– |
Yes / Yes / Yes |
| Connector type |
– |
3 m 7-pin LEMO |
| Weight |
g / oz |
1550 / 54.675 |
.jpg)
Fig 1. The transfer impedance of a typical RA0403 Hi-Res Ear Simulator with tolerances. Up to 10 kHz, the response is within the limits defined by IEC 60318-4. Above 10 kHz, the response is within a fairly tight tolerance band up to 50 kHz