The GRAS 43AC-S1 CCP Ear Simulator Kit according to IEC 60318-4 with TEDS is a complete test-jig for acoustical measurements on earphones coupled to the ear by inserts such as tubes and ear moulds.
Technology
The GRAS 43AC-S1 is a complete test jig for acoustically testing earphones coupled to the ear via inserts such as tubes and ear moulds and complies with the following international requirements:
- IEC 60318-4: Occluded-ear simulator for the measurement of earphones coupled to the ear by ear inserts.
The 43AC-S1 comprises:
- GRAS RA0045-S1 Ear Simulator
- GRAS 40AO 1/2" Prepolarized Pressure Microphone
- GRAS 26CB 1/4" CCP Standard Preamplifier with Microdot Connector
- GRAS RA0001 Right-angled Adapter
- GRAS RA0052 Test Jig
- GRAS AA0070 3m Microdot-BNC Cable
The Test Jig has an adjustable spring-loaded arm to exert a variable force on the test object.
Click GRAS RA0045-S1 for more information about the RA0045-S1 Ear Simulator.
An externally polarized version is available, GRAS 43AC.
TEDS Compatibility
43AC-S1 is IEEE 1451.4 TEDS v. 1.0 compliant. If your measurement platform supports Transducer Electronic Data Sheets (TEDS), you will be able to read and write data like properties and calibration data.
Specifications
| Parameter |
Unit |
Value |
| Theoretical lower dynamic range |
dB(A) |
25 |
| Theoretical upper dynamic range |
dB |
150 |
| Sensitivity @ 250 Hz (±2 dB) |
mV/Pa |
12 |
| Sensitivity @ 250 Hz (±2 dB) |
dB re 1 V/Pa |
-38.5 |
| Coupler volume |
mm³ |
1260 @ 500 Hz |
| Resonance frequency |
kHz |
13.5 |
| Operating temperature range |
°C / °F |
-30 – 60 / -22 – 140 |
| Temperature coefficient @ 250 Hz |
dB/°C / dB/°F |
-0.01 / -0.006 |
| Non-condensing humidity range |
% RH |
0 – 75 |
| ANSI standard |
|
S3.7 |
| IEC standard |
|
60318-4 (formerly 60711) |
| CE / RoHS / WEEE compliance |
|
Yes / Yes / Yes |
| Connector type |
|
Microdot 10/32 |
| Weight |
g / oz |