GRAS 43AA-S2 CCP Ear Simulator Kit According to IEC 60318-1 & -2

  •  Part Number: GRAS 43AA-S2
     Category: GRAS Sound & Vibration
  •  View: 8
     Status: Available
  • Connection: 0 V/CCP
    Volume: Complex
    Dyn range: 25 dB(A) to 150 dB
    ANSI: S3.7
    IEC: 60318-1 & -2

  • Sale Price: Liên hệ
Product Information

The GRAS 43AA-S2 CCP Ear Simulator Kit with TEDSis a complete test jig for acoustical measurements on telephone handsets and earphones. It includes an IEC 60318 Ear Simulator RA0039 which is individually calibrated according to the ITU-T Recommendation P.57 (08/96) ”Series P: Telephone transmission quality, objective measuring apparatus: artificial ears.

Technology

The GRAS 43AA-S2 CCP Ear Simulator Kit according to IEC 60318-1 & -2 is a complete test jig for acoustical measurements on telephone handsets and earphones in accordance with:

  • IEC60318-1 & -2 Electroacoustics – Simulators of human head and ear – Part 1: Ear simulator for the calibration of supraaural earphones, 1998-07
  • ITU-T Recommendation P.57 (08/96) Series P: Telephone transmission Quality, Objective measuring apparatus: Artificial Ears

It consists of the following components:

  • IEC 60318-1 & -2 GRAS RA0039 Ear Simulator 
  • GRAS 40AO ½" Pressure Microphone 
  • GRAS 26CB ¼" Preamplifier 
  • GRAS RA0052 Test Jig
  • Mounting plates for circum-aural and supra-aural headphones

The Test Jig has an adjustable spring-loaded arm to exert a variable force on the test object.

An externally polarized version is available, GRAS 43AA.

A version with a GRAS 26AB. Preamplifier is available, GRAS 43AA-S3.

TEDS Compatibility

43AA-S2 is IEEE 1451.4 TEDS v. 1.0 compliant. If your measurement platform supports Transducer Electronic Data Sheets (TEDS), you will be able to read and write data like properties and calibration data.

Specifications

Parameter Unit Value
Theoretical lower dynamic range dB(A) 25
Theoretical upper dynamic range dB 150
Sensitivity @ 250 Hz (±3 dB) mV/Pa 12
Sensitivity @ 250 Hz (±3 dB) dB re 1 V/Pa -38.5
Operating temperature range °C / °F -30 – 60 / -22 – 140
Temperature coefficient @ 250 Hz dB/°C / dB/°F -0.01 / -0.006
Non-condensing humidity range % RH 0 – 80
ANSI standard   S3.7
ITU-T recommendation   Page 57, Type 1
CE / RoHS / WEEE compliance   Yes / Yes / Yes
Connector type   BNC
Weight g / oz 1.65 / 0.058